10.4.2 Assemble electronics housing
Install EMI filter and terminal block
Procedure
1. Install replacement EMI filter (D, Figure 10-26) and/or terminal block (F) on
electronic stack.
2. Refer to wiring details in Figure 10-26. Connect EMI filter and ground wire
(E) at terminal block (F).
3. Connect EMI filter wiring at AC power input terminal block on electronic
stack.
Install solenoid valves
Procedure
1. Disassemble replacement solenoid valve (I or J, Figure 10-25).
2. Install new solenoid valve base.
Be careful not to overtighten.
3. Install new washer and solenoid coil assembly and secure with nut.
4. Connect the solenoid leads to the proper terminations on the solenoid
power terminal block (Figure 10-27).
Maintenance and service Manual
December 2022 00809-0500-4880
284 Emerson.com/Rosemount