4. Replace the cover.
The two tabs at the back edge of the cover fit into slots at the rear of the enclosure,
and the three small slots in the front of the cover snap into the three tabs next to
the relay terminal strip. See Alarm relay connection.
5. Once the tabs are lined up, push the cover to snap it into place.
6. Run the power wiring through the conduit opening nearest the power terminal.
AC power wiring should be 14 gauge or greater.
7. Provide a switch or breaker to disconnect the transmitter from the main power
supply.
8. Install the switch or breaker near the transmitter and label it as the disconnecting
device for the transmitter.
4.4  Wire analog outputs
Four analog current outputs are located on the main circuit board, which is attached to the
inside of the enclosure door.
Figure 4-2 shows the locations of the terminals, the outputs they are assigned to, and the
polarity.
Reference Manual Wiring
00809-0200-3415 January2023
Rosemount TCL 56 19