Technical data Cerabar PMC11, PMC21, PMP11, PMP21, PMP23
42 Endress+Hauser
Device Connection Climate class Option
1)
PMC11
PMP11
PMC21
PMP21
PMP23
Valve plug ISO4400 M16 IP65 NEMA type 4X enclosure U
PMC11
PMP11
PMC21
PMP21
PMP23
Valve plug ISO4400 NPT ½ IP65 NEMA type 4X enclosure V
1) Product Configurator, order code for "Electrical connection"
2) IP 68 (1.83m H2O for 24 h)
3) Designation of the IP protection class according to DIN EN 60529. Previous designation "IP69K" according
to DIN 40050 Part 9 is no longer valid (standard withdrawn on November 1, 2012). The tests required by
both standards are identical.
12.5.5 Vibration resistance
Test standard Vibration resistance
IEC 60068-2-64:2008 Guaranteed for 5 to 2000Hz: 0.05g
2
/Hz
12.5.6 Electromagnetic compatibility
Interference emission as per EN 61326-1 equipment B
Interference immunity as per EN 61326-1 (industrial environment)
NAMUR recommendation EMC (NE21)
Maximum deviation: 1.5% for TD 1:1
For more details, refer to the Declaration of Conformity.