EasyManua.ls Logo

Endress+Hauser Ecograph T RSG30 - Page 49

Endress+Hauser Ecograph T RSG30
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning
49
6.4.3 Set up - Signal analysis
Function only visible if a digital input has been activated as a counter, or the option "Integration +
Analysis + Mathematic" is activated!
Set up in order to get signal analysis for a preset time range/cycle as well as the function to manually
reset the signal analysis.
The analyses can be displayed in normal operation at the touch of a button ("Main menu - Analysis").
!
Note!
Signal analysis means a quantity analysis or an operational time analysis (built-in function) and a
min/max/average value analysis (with option "integration + analysis + mathematic") within the
adjusted period.
This information is also saved (reduces the memory available for the graphics), transferred to a PC
and can be used there for further analyses.
Fig. 34: Set up, Signal analysis
"Signal analysis"
menu items
Configurable parameters
(Factory settings bold marked)
Intermed analysis Calculates quantities and operating times for the time cycles preset here. Picklist: No, External, 1min,...12h
Day Calculates daily quantities and operation times. Picklist: no (off), yes (on)
Month Calculates monthly quantities and operation times. Picklist: no (off), yes (on)
Total/year count. Calculates quantities and operation times.
"Yes": analysis time cycle is one year.
"No(=Total)": the analysis is continuous from the time of the last reset (useful as a totalizer).
Synchron. time Time for ending the signal analysis. If, for example, 07:00 is set up then the daily analysis will run from 07:00 of the actual day until
07:00 of the following day. (00:00)
Reset to zero The analyzed values can be reset to zero. Example: Reset after commissioning plant. All (installation) signals are deleted. The graphics
or memory storage are not influenced (traceability)!
Picklist: no, intermed analysis, daily counter, monthly counter, total/yearly counter, all counters.
!
Note!
All previous (commissioning) signals are discarded.
If the previous signals are still required, save them on the CompactFlash card before resetting (see Section "Commissioning -
Accessible functions - CompactFlash card").
The reset takes effect immediately if you confirm here with "E= Accept".

Table of Contents