115 (4.53)
20.5 (0.8)
110 (4.3)
15.5 (0.6)
Technical data iTEMP TMT71/72
TA30H with display window in cover Specification
125 (4.92)
28
(1.1)
78 (3.01)
Flameproof (XP) version, explosion-protected, captive screw
cap, with two cable entries
Temperature: –50 to +150 °C (–58 to +302 °F) for rubber seal
without cable gland (observe max. permitted temperature of
cable gland!)
Material:
Aluminum with polyester powder coating
Stainless steel 316L without coating
Cable entry glands: ½" NPT, M20x1.5
Color of aluminum head: blue, RAL 5012
Color of aluminum cap: gray, RAL 7035
Weight:
Aluminum: approx. 860 g (30.33 oz)
Stainless steel: approx. 2 900 g (102.3 oz)
A0009831
TA30D Specification
107.5 (4.23)
2 cable entries
Temperature: –50 to +150 °C (–58 to +302 °F) without cable
gland
Material: aluminum, polyester powder coated
Seals: silicone
Cable entry glands: 1/2" NPT and M20x1.5
Two head transmitters can be mounted. In the standard
configuration one transmitter is mounted in the terminal
head cover and an additional terminal block is installed
directly on the insert.
Color of head: blue, RAL 5012
Color of cap: gray, RAL 7035
Weight: 390 g (13.75 oz)
28
(1.1)
78 (3.1)
A0009822
Weight Head transmitter: approx. 40 to 50 g (1.4 to 1.8 oz)
Field housing: see specifications
Materials All the materials used are RoHS-compliant.
Housing: polycarbonate (PC), corresponds to UL94, V-2 UL recognized
Terminals:
Screw terminals: nickel-plated brass and gold-plated contacts
Potting (head transmitter): QSIL 553
Field housing: see specifications
13.7 Certificates and approvals
CE mark The product meets the requirements of the harmonized European standards. As such, it
complies with the legal specifications of the EC directives. The manufacturer confirms
successful testing of the product by affixing to it the CE-mark.
58 V. 1, Rev. 2, 4-10-2018 Endress+Hauser