EasyManua.ls Logo

Endress+Hauser Micropilot FMR67 - Page 38

Endress+Hauser Micropilot FMR67
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR67
38 V. 1, Rev. 3, 11-04-2017 Endress+Hauser
FMR67, flush mount antenna, PTFE DN80, standard flange 316L
p
-40
(-40)
+150
(+302)
+200
(+392)
0
(+32)
T
p
[°C]
([°F])
1
2
[bar] ([psi])
16 (232)
-1 (-14.5)
0 (0)
A0032187
14 FMR67: Permitted range for process temperature and process pressure, antenna, PTFE DN80, standard
flange 316L
1 Feature 90, seal: A5, FKM Viton GLT
2 Feature 90, seal: A6, FKM Viton GLT
FMR67, PTFE DN80, standard flange 316L
Feature 100 "Process connection" Feature 90
"Seal"
Process temperature
range
Process pressure range
AGJ:
NPS 3" Cl.150 RF, 316/316L
AHJ:
NPS 4" Cl.150 RF, 316/316L
CGJ:
DN80 PN10/16 B1, 316L
CHJ:
DN100 PN10/16 B1, 316L
KGJ:
10K 80A RF, 316L
KHJ:
10K 100A RF, 316L
A5, FKM Viton
GLT
–40 to +150 °C
(–40 to +302 °F)
p
rel
=
–1 to 16 bar
(–14.5 to 232 psi)
1)
A6, FKM Viton
GLT
–40 to +200 °C
(–40 to +392 °F)
1) The pressure range may be further restricted in the event of a CRN approval
FMR67, flush mount antenna, PTFE DN80, UNI flange 316L
p
-40
(-40)
+150
(+302)
+200
(+392)
0
(+32)
T
p
[°C]
([°F])
1
2
[bar] ([psi])
1 (14.5)
-1 (-14.5)
0 (0)
A0032199
15 FMR67: Permitted range for process temperature and process pressure, antenna, PTFE DN80, UNI flange
316L
1 Feature 90, seal: A5, FKM Viton GLT
2 Feature 90, seal: A6, FKM Viton GLT

Table of Contents