32
Nivotester FTR525
Endress+Hauser
Commissioning
8�4�2 Output adjustments
The following functions are used to adjust the signal outputs, among others the
assignment and the behavior of the relay/SSR and current outputs.
Example: Backlog monitoring
material transport is detected by the transceiver FDR57. In combination with the signal of
3
FQR57 FDR57
FQR57 FDR57
FQR57 FDR57
22 Exampleofcombinedlimitandmotiondetection
In this example featuring relay outputs, switching contact 1 (2210) is assigned to the limit
level detection (1M) and switching contact 2 (2220
(2M), minimum safety was chosen for both switching functions (2411+2421).
This leads to the following conditions at the signal outputs:
Pos� Condition Limit Switch.outp.1
(3 NO - 4 CC)
Bulkow Switch.outp.2
(15 NO - 16 CC)
1
Continuous
2
3 Backlog