Nivotester FTR525
Endress+Hauser
41
Commissioning
8�6�1 Simulation mode
Diagnose
1000
→
Simulation
1700
→
Simulation mode
1710
values and signal outputs).
"Current output", "Switching contact 1", "Switching contact 2", "Failure contact"
8�6�2 Simulation value
Diagnose
1000
→
Simulation
1700
→
Simulation value
1720
This function is used to enter corresponding simulation values for the selected simulation
Options: Measurement limit 0 to 100 %, measurement motion 0 to 100 %,
current output 3.6 to 22.0 mA, contact 3-4 opened, contact 3-4 closed,
contact 15-16 opened, contact 15-16 closed, failure, no failure
Simulation Simulation
value
Beschreibung
Measurement
limit 0-100%
Measurement
limit 0 to 100 %
• A corresponding measurement is simulated by press-
• All signal outputs follow the simulated value accord-
ing to the settings.
Measurement
motion 0-100%
Measurement
motion 0 to
100 %
Current output Current output
3�6 to 22�0 mA
• A corresponding current output value is simulated by
• All other outputs follow the original settings.
Switching
contact 1
• Contact 3-4
opened
• Contact 3-4
closed
• Contact 3-4 of switching contact 1 is alternately
• All other outputs follow the original settings.
Switching
contact 2
• Contactt 15-16
opened
• Contact 15-16
closed
• Contact 15-16 of switching contact 2 is alternately
• All other outputs follow the original settings.
• This simulation is only available on FTR525 with two
switching outputs (relay or SSR).
Failure contact • No failure
• Failure
• The failure contact (C+/C-) is alternately opened or
• All other outputs follow the original settings.
O • The simulation is deactivated.
• All other outputs follow the original settings.