Commissioning
8.4 Conguringthemeasuringdevice
well as the behavior of the display (such as the operating language and display mode).
• The numbers assigned to the functions and function groups are added in brackets for
easier navigation in the following sub-sections.
• Overview of all available functions and factory settings
8�4�1 Adjustments to the application
The Soliwave FQR57/FDR57 microwave barrier must be adjusted to suit the application
for commissioning; the following functions of the application main function group (2000)
can be used for this.
Details of the Soliwave FQR57/FDR57 microwave barrier: Technical Information
TI01330F/97/EN and Operating Instructions BA01804F/97/A2
8�4�1�1 Automatic setup of level limit detection
Application
2000
→
Sensor
2100
→
Limit detection
2110
→
Automatic setup
2111
Automatic setup is carried out with either a free (1) or a covered (2) path; the option
selected for this corresponds to the state of the path.
Options: "Free path", "Covered path"
16 Setupconditionsofthelimitdetection
the microwave barrier or to suit changing application conditions.
8�4�1�2 Manualadjustmentlowerlimit(LL)levellimitdetection
Application
2000
→
Sensor
2100
→
Limit detection
2110
→
Manual adjustment LL
2112
range 0 to 100 %).
To assist with adjustment, the current measured value is shown alongside the set
detection range.