Proline Promag 50H, 53H
Endress+Hauser 41
Clamp DIN 32676 Sensor Fits to di G D L LK
1.4404 / 316L
5*H**-0***********
DN
[mm]
Piping
DIN 11850 [mm] [mm] [mm] [mm] [mm]
A0005539
40 42 × 2 38 50.5 92 61.5 71.0
50 54 × 2 50 64.0 105 61.5 83.5
65 70 × 2 66 91.0 121 68.0 100.0
80 85 × 2 81 106.0 147 68.0 121.0
100 104 × 2 100 119.0 168 68.0 141.5
– Fitting length for DN 40 to 65 = (2 × L) + 136 mm
– Fitting length for DN 80 to 100 = (2 × L) + 196 mm
If pigs are used for cleaning, it is essential to take the inside diameters of
measuring tube and process connection (di) into account!
Tri-Clamp L14 AM7 Sensor Fits to di G D L LK
1.4404 / 316L
5*H**-1***********
DN
[mm]
DN
[inch]
Piping
OD [mm] [mm] [mm] [mm] [mm]
A0005543
40 1½" 38.1 × 1.65 34.8 50.4 92 68.8 71.0
50 2" 50.8 × 1.65 47.5 63.9 105 68.8 83.5
65 63.5 × 1.65 60.2 77.4 121 68.8 100.0
80 3" 76.2 × 1.65 72.9 90.9 147 68.8 121.0
100 4" 101.6 × 1.65 97.4 118.9 168 68.8 141.5
– Fitting length for DN 40 to 65 = (2 × L) + 136 mm
– Fitting length for DN 80 to 100 = (2 × L) + 196 mm
If pigs are used for cleaning, it is essential to take the inside diameters of
measuring tube and process connection (di) into account!
Coupling SC DIN 11851 Sensor Fits to di G D L LK
1.4404 / 316L
5*H**-2***********
DN
[mm]
Piping
DN 11850 [mm] [mm] [mm] [mm] [mm]
A0005540
40 42 × 2 38 Rd 65 × 1/6" 92 72 71.0
50 54 × 2 50 Rd 78 × 1/6" 105 74 83.5
65 70 × 2 66 Rd 95 × 1/6" 121 78 100.0
80 85 × 2 81 Rd 110 × 1/6" 147 83 121.0
100 104 × 2 100 Rd 130 × 1/6" 168 92 141.5
– Fitting length for DN 40 to 65 = (2 × L) + 136 mm
– Fitting length for DN 80 to 100 = (2 × L) + 196 mm
If pigs are used for cleaning, it is essential to take the inside diameters of
measuring tube and process connection (di) into account!
Coupling DIN 11864-1 Sensor Fits to di G D L LK
Aseptic threaded adapter, Form A
1.4404 / 316L
5*H**-3***********
DN
[mm]
Piping
DN 11850 [mm] [mm] [mm] [mm] [mm]
A0005545
40 42 × 2 38 Rd 65 × 1/6" 92 71 71.0
50 54 × 2 50 Rd 78 × 1/6" 105 71 83.5
65 70 × 2 66 Rd 95 × 1/6" 121 76 100.0
80 85 × 2 81 Rd 110 × 1/6" 147 82 121.0
100 104 × 2 100 Rd 130 × 1/6" 168 90 141.5
– Fitting length for DN 40 to 65 = (2 × L) + 136 mm
– Fitting length for DN 80 to 100 = (2 × L) + 196 mm
If pigs are used for cleaning, it is essential to take the inside diameters of
measuring tube and process connection (di) into account!
G
di
LK
D
L
G
di
LK
D
L
L
G
di
D
LK
L
G
di
D
LK

Table of Contents