Installation Proline Promass F 100 Modbus RS485
24
NOTICE
Electronics overheating on account of thermal insulation!
Recommended orientation: horizontal orientation, transmitter housing pointing
downwards.
Do not insulate the transmitter housing .
Maximum permissible temperature at the lower end of the transmitter housing:
80 °C (176 °F)
Thermal insulation with extended neck free: the insulation is omitted around the
extended neck. We recommend that you do not insulate the extended neck in order to
ensure optimum dissipation of heat.
A0034391
 7 Thermal insulation with extended neck free
Heating
NOTICE
Electronics can overheat due to elevated ambient temperature!
Observe maximum permitted ambient temperature for the transmitter .
Depending on the fluid temperature, take the device orientation requirements into
account .
NOTICE
Danger of overheating when heating
Ensure that the temperature at the lower end of the transmitter housing does not
exceed 80 °C (176 °F).
Ensure that convection takes place on a sufficiently large scale at the transmitter neck.
Ensure that a sufficiently large area of the housing support remains exposed. The
uncovered part serves as a radiator and protects the electronics from overheating and
excessive cooling.
Heating options
If a fluid requires that no heat loss should occur at the sensor, users can avail of the
following heating options:
Electrical heating, e.g. with electric band heaters
Via pipes carrying hot water or steam
Via heating jackets
Using an electrical trace heating system
If heating is regulated via phase angle control or pulse packages, magnetic fields can affect
the measured values (= for values that are greater than the values permitted by the EN
standard (sine 30 A/m)).
For this reason, the sensor must be magnetically shielded: the housing can be shielded
with tin plates or electric sheets without a privileged direction (e.g. V330-35A).