EasyManua.ls Logo

Endress+Hauser Proline Promass F 100 - Diagnostic Information via Light Emitting Diodes; Diagnostic Information in DeviceCare or FieldCare

Endress+Hauser Proline Promass F 100
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promass F 100 Modbus RS485 Diagnostics and troubleshooting
77
Error Possible causes Solution
Operation with FieldCare or
DeviceCare via CDI-RJ45 service
interface (port 8000)
Firewall of computer or network is
preventing communication
Depending on the settings of the
firewall used on the computer or in
the network, the firewall must be
adapted or disabled to allow
FieldCare/DeviceCare access.
Flashing of firmware with FieldCare
or DeviceCare via CDI-RJ45 service
interface (via port 8000 or TFTP
ports)
Firewall of computer or network is
preventing communication
Depending on the settings of the
firewall used on the computer or in
the network, the firewall must be
adapted or disabled to allow
FieldCare/DeviceCare access.
12.2 Diagnostic information via light emitting diodes
12.2.1 Transmitter
Different LEDs in the transmitter provide information on the device status.
LED Color Meaning
Supply voltage Off Supply voltage is off or too low
Green Supply voltage is ok
Alarm Off Device status is ok
Flashing red A device error of diagnostic behavior "Warning" has occurred
Red A device error of diagnostic behavior "Alarm" has occurred
Boot loader is active
Communication Flashing white Modbus RS485 communication is active
12.2.2 Safety Barrier Promass 100
Various light emitting diodes (LEDs) on the Safety Barrier Promass 100 provide status
information.
LED Color Color
Power Off Supply voltage is off or too low.
Green Supply voltage is ok.
Communication Flashing white Modbus RS485 communication is active.
12.3 Diagnostic information in DeviceCare or FieldCare
12.3.1 Diagnostic options
Any faults detected by the measuring device are displayed on the home page of the
operating tool once the connection has been established.

Related product manuals