Proline Prosonic Flow P 500
44 Endress+Hauser
1
2
3
500 mm (20 in)
3
A0043374
 32 Holder with strapping bands (large nominal diameters)
1 Strap bolt with guide*
2 Strapping band*
3 Tensioning screw
*The distance between the strap bolt and strapping band lock must be at least 500 mm (20 in).
For 1-traverse mounting with 180° (opposite) →  7 (single-path measurement,
A0044304), →  6,  8(two-path measurement, A0043168)
For 2-traverse mounting →  7 (single-path measurement, A0044305), →  7,  8(two-
path measurement, A0043309)
Electrical connection →  8,  10
Sensor holder with welded bolts
Can be used for
Measuring devices with measuring range DN 50 to 4000 (2 to 160")
Mounting on pipes DN 50 to 4000 (2 to 160")
Procedure:
The welded bolts must be fixed at the same installation distances as the mounting bolts with
strapping bands. The following sections explain how to align the mounting bolts, depending on
the mounting method and measurement method:
Installation for measurement via 1 traverse →  47
Installation for measurement via 2 traverses →  49
The sensor holder is secured as standard with a locking nut with a metric M6 ISO thread. If
another thread should be used for fastening purposes, a sensor holder with a detachable locking
nut must be used.