11 Description of device functions Prosonic Flow 92
82 Endress+Hauser
System Reset Use this function to re-initialise the memory and reset all current settings to
the factory settings.
Options:
NO – YES
Caution!
During memory initialisation, all data sets including Site Setup and log data
are erased irreversibly.
Exception:
The operating language selection remains unchanged as well as the settings
of the following functions:
Pipe Diameter (see Page 69)
Pipe Material (see Page 69)
Wall Thickness (see Page 70)
Liner Material (see Page 70)
Liquid (see Page 70)
Sensor Config. (see Page 71)
Sensor Type (see Page 71)
Burst Voltage (see Page 71)
Input Range (see Page 84)
Unit (Output) (see Page 85)
Output Range (see Page 85)