Prosonic S FMU90 PROFIBUS DP Installation
Endress+Hauser 17
5.3.1 Installation conditions
Dimensions
92 (3.62)
104 (4.09)
143 (5.63)
43 (1.69)
EN 60715
TH 35x7.5/15
(1.4x0.3/0.6)
35 (1.38)
150 (5.91)
104 (4.09)
143 (5.63)
EN 60715
TH 35x7.5/15
(1.4x0.3/0.6)
35 (1.38)
43 (1.69)
208 (8.19)
143 (5.63)
EN 60715
TH 35x7.5/15
(1.4x0.3/0.6)
35 (1.38)
43 (1.69)
104 (4.09)
Σ = 0 Σ = 1...3
Σ = 4
A0035915
 11 Dimensions of Prosonic S with DIN rail housing; Σ: number of additional connection modules. Unit of
measurement mm (in)
Mounting location
In cabinet outside hazardous areas
At a sufficient distance from high-voltage electric cables, motor cables, contactors or
frequency converters
Altitude: install at a maximum altitude of 2 000 m (6 560 ft) over MSL
Minimum clearance to the left: 10 mm (0.4 in); the housing cover cannot be opened
otherwise.