Prosonic S FMU90 PROFIBUS DP Electrical connection
Endress+Hauser 21
6 Electrical connection
6.1 Connection conditions
6.1.1 Cable specification
Conductor cross-section: 0.2 to 2.5 mm
2
(26 to 14 AWG)
Wire sleeve cross-section: 0.25 to 2.5 mm
2
(24 to 14 AWG)
Min. stripping length: 10 mm (0.39 in)
6.2 Connecting the device
6.2.1 Terminal compartment of polycarbonate field housing
Valid for:
Order code 030 (housing, material)
Option 1 (PC field mounting, IP66 NEMA4x)
Access to terminal compartment
1.
2.
3.
A0034895
 16 Access to terminal compartment in polycarbonate field housing
1.
2.
3.
A0034896
 17 For easier wiring, remove the cover from the field housing