 58 Configuration of a time pulse relay
A Pulse time
B Pulse width
C Relay energized
D Relay de-energized
Navigation
Relay/controls → Relay configurat. → Relay N
Parameter
Select function
Select the Time pulse option.
Pulse width
Define the duration of each individual pulse.
Pulse time
Define the time interval between the individual pulses.
Invert
Specify whether the switching direction of the relay should be inverted. (Factory setting:
No).
Error handling
Specify the behavior of the relay if an error occurs.
9.4.8 Configuring the counting pulse relay
A counting pulse relay generates a short pulse each time a certain volume of liquid has
flowed through the flume or weir.
Navigation
Relay/controls → Relay configurat. → Relay N
Parameter
Select function
Select the Counting pulse option.
Function
Select the flow to which the counting pulses refer.
Counter unit
Select the volume unit for flow counting.
Pulse value
Select the flow volume after which a pulse is generated.
Pulse width
Define the duration of each individual pulse.
Pulse counter
Indicates how many pulses have already been generated.
Overflow x 10^7
Indicates how often the counter has passed the overflow point of 10
7
counting pulses.
Reset counter
Yes: The counter is reset to 0.
No: The counter keeps its value.
Start counter
Specify the lower flow limit for counting. Flows below this value are ignored during
counting.