Operating Instructions Raman Rxn2
18 Endress+Hauser
Figure 5. Cleaning electro-optical fiber connection
3. Next, use an IBC 1.25 mm ferrule cleaner with the bulkhead adapter attached to do a final clean of the center of
the ferrule where the fiber resides. Press together until a click is heard and repeat once.
Figure 6. Final clean of electro-optical fiber connector fiber tips
4. Release the latch and open the spring-loaded cap on the Raman Rxn2 analyzer device-side connector. Insert the
cable-side connector into the instrument device-side connector fully and engage latch to secure. The connectors
are polarized and can only be inserted one way. The slotted head screws on the face of both connectors should be
facing out.
Figure 7. Connecting an electro-optical fiber cable to a Raman Rxn2 four-channel analyzer
5. Repeat for each probe.