12.1.3 Laser
Item
Description
532 nm
Excitation wavelength
Maximum power output
Warranty
532 nm
120 mW
1 year or 5000 hours
785 nm
Excitation wavelength
Maximum power output
Warranty
785 nm
400 mW
unlimited hours for 1 year
993 nm
Excitation wavelength
Maximum power output
Warranty
993 nm
400 mW
unlimited hours for 1 year
12.1.4 Probes
Analyzer configuration
Probe compatibility
Raman Rxn2 single-channel, four-channel,
and starter
Compatible with:
Rxn-10 probe equipped with immersion or non-contact optics
Endress+Hauser liquid phase Raman probes
Endress+Hauser bioprocessing Raman probes
Raman Rxn2 hybrid
Compatible with:
Rxn-20 probe and 1 other ALT probe including:
Rxn-10 probe equipped with immersion or non-contact optics
Endress+Hauser liquid phase Raman probes
Endress+Hauser bioprocessing Raman probes
12.2 Certifications
Raman Rxn2 analyzers carry certifications for installation in a general-purpose zone with output into hazardous
areas. For more specific information on the hazardous area rating for field measurements, refer to the Operating
Instructions for the installed probe.
Certification: base unit (fiber-optic and interlock outputs only)
Certification
Marking
Temperature (ambient)
IECEx
Ex [ia Ga] [op sh Gb] IIC
5 to 35 °C (41 to 95 °F)
ATEX
II (2)(1) G
Ex [ia Ga] [op sh Gb] IIC
5 to 35 °C (41 to 95 °F)
North America
Class I, Division 1, Groups A, B, C and D or [Ex ia] Class I, Division 1,
Groups A, B, C, and D:
[Ex ia Ga] IIC Class I, Division 2, Groups A, B, C and D:
[Ex ia Ga] [op sh Gb] IIC
5 to 35 °C (41 to 95 °F)
UKCA
II (2)(1) G
Ex [ia Ga] [op sh Gb] IIC
5 to 35 °C (41 to 95 °F)
JPEx
Ex [ia Ga] [op sh Gb] IIC
5 to 35 °C (41 to 95 °F)