Raman Rxn2 Operating Instructions
Endress+Hauser 23
6 Electrical and I/O connections
6.1 Overview of rear panel electrical and I/O connections
Figure 16. Rear panel of Raman Rxn2 four-channel analyzer
Remote interlock connectors
Safety feature. To interrupt the laser, remove the black plug.
Provides fiber-optic laser radiation output, fiber-optic Raman collection and
electrical laser interlock loop for each instrument channel. The electrical laser
interlock loop is intrinsically safe and is governed by Endress+Hauser drawing
4002396.
Match 3 prongs on probe to 3 plugs on EO. Pull latch down to secure probe in place.
Laser radiation may NOT exit a channel whose EO fiber connector is removed
because removing the EO connector also interrupts the laser interlock loop for that
channel.
Air exhaust fan and outlet.
Product information about the Raman Rxn2 analyzer.
Touchscreen USB port, USB port, Ethernet ports, RS-485 serial port, and touchscreen
video port.
AC inlet AC 100 to 240 V 50/60 Hz
Power socket outlet that provides AC power to the base unit. Ground pin on this
connector serves as the protective conductor terminal.
EO fiber cable strain relief mounting location.