9.2 Raman Rxn2 system and power loss
The instrument retains its last known power state in non-volatile memory. Should power be interrupted to the
instrument at any time, the instrument will recall its last known power state and return to that power state once
power is restored. For example, if the instrument was ON when power was interrupted, it will power on
automatically once power is restored. If the laser was ON and the laser key is also in the ON position as well, then the
laser will turn on. In the unlikely event of this occurring, this represents a possible laser exposure hazard. While
power is interrupted, if the instrument was ON at the time of power loss, the power switch will flash a two-blink
error code for 30 to 60 seconds, indicating that power has been lost.