Operating Instructions Raman Rxn2
20 Endress+Hauser
Figure 9. Employing strain reliefs to secure the fiber assembly for the Rxn-20 channel
4. Remove the threaded cap from the calibration (CAL) port.
5. Connect the clean calibration fiber to the CAL port, aligning the key on the fiber connector with the notch on the
CAL port connector. The fiber should be routed in a helical pattern as shown below to avoid kinks.
Figure 10. Proper connection and routing of the calibration fiber
6. Remove the rubber cap from the MT style collection fiber connector. Lift and hold the cover of the collection
(COLL) port as the fiber is connected. Align the white dot on the MT style fiber connector with the white mark on
the COLL port and insert the fiber connector until it clicks into place. The fiber should be routed in a helical
pattern as shown below to avoid kinks.
Figure 11. Proper connection and routing of the collection fiber
7. Remove the excitation fiber clamp by loosening the clamp screw with a 3/32" ball driver (provided with analyzer)
and sliding it out of the way.
8. Remove the threaded cap from the excitation (EXC) port.
A0048446
A0048453
A0048454