Operating Instructions Raman Rxn2
26 Endress+Hauser
Figure 18. Rear panel of Raman Rxn2 hybrid configuration with Rxn-20 channel cover attached (left) and removed (right)
7.2.1 Power and grounding
The Raman Rxn2 has a standard IEC-320 C-14 inlet for power at the rear of the instrument. Any power cord with an
IEC-320 C-13 plug will connect to the base unit. The Raman Rxn2 accepts AC power from 100 to 240 V and 50/60
Hz. For U.S. applications, a power cord is supplied. For non-U.S. applications, the user must supply a power cable that
meets local/national standards.
Also included on the rear of the instrument is a functional earth terminal for additional grounding if required.
Primary grounding occurs through the IEC power inlet plug ground terminal which should be connected to the
building grounding system.
Do not position the Raman Rxn2 in a way that makes it difficult to remove the mains cord. Only use adequately rated
power cables with the Raman Rxn2 system.