Parameter Setting Up selections
Default
settings
Actual
Numeric display
brightness
BRNUM
Increase or decrease 7
segment display
brightness
Value:
0to9(9=max.)
5
Bargraph brightness
BRLED
Increase or decrease the
bargraph and set point
diode brightness
Value:
0 to 9 (9 = max.)
5
Programme name
PNAME
Display address:
Display of the software
used in the unit.
Software version
SU-ID
Display address:
Display of the version
number of the software
used
Power frequency
FREQ
Power supply frequency;
This is required in order
to eliminate any super-
imposed interferance of
the power supply
frequency on the signal
measurement.
50 HZ 50 Hz power supply
60 HZ 60 Hz power supply
50 HZ
Test
TEST
Test function for diverse
hardware components;
These are activated after
the component is
selected.
OFF None
REL1 Relay 1 is energised
REL2 Relay 2 is energised
DISP All segments on the
numberic display and
all LEDs are activated
for approx. 5 s.
OFF
Energised relay means
rest position, in fault alarm
and set point infringement
the relay is de-energised.
Once this address has been exited it is automatically reset to
OFF.