EasyManua.ls Logo

Endress+Hauser RMA 422 - 8. Technical Data (Continued); Display, Certification, and Approval Specifications

Endress+Hauser RMA 422
150 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Output Number max. 2
Fault condition 3.5 mA or 22 mA presettable
Effect to NAMUR recommendation NE43
Response time max. 200 ms (input signal from 10% to 90% FSD)
D/A resolution Current: 13 bit, Voltage: 13 bit
Number max. 2
Galvan. isolation To all other current circuits
Relays
Output signal Binary, switches on reaching a preset alarm set point value
Number 2
Contact type 1 potential free changeover contact (SPDT)
Contact loading <= 250 V
AC
,5A/30V
DC
,5A
Alarm set point function
Operating mode Off, min-, Maximum safety, gradient, alarm
Switch threshold - 19999 to + 99999
Hysteresis - 19999 to + 99999
Time delay 0s to 99s
Signal source Input 1, input 2, mathematics process value
Number 2
Display 1 yellow LED per set point, optional symbol in LC display
Scan rate 100 ms
Mathematics function
Operator Addition/subtraction/multiplication
Factor 1/2 -19999 to +99999, 0 to 4 decimal points
Input 1/2 Dependent on selected input values
Offset -19999 to +99999, 0 to 4 decimal points
Power supply Power supply 90...250 V
AC
50/60 Hz
18...36 V
DC
, 20...28 V
AC
50/60 Hz
Power
consumption
11 VA
Fuse 315 mA, slow blow (90...250 V) 630 mA, slow blow (20...28 V)
Accuracy Current Accuracy: 0.1 % FSD
Temperature drift: 0.05 % / 10 K ambient temperature
Analogue output Accuracy: 0.1 % FSD
Temperature drift: 0.05 % / 10 K ambient temperature
Process transmitter
70

Table of Contents