5.3 Display/measurement range
Parameter Setting up selections
Default
settings
Actual
setting
* Reference value
DISOW
Selection of signal source of
the display value
Selection range:
IN1: Process value ch. 1
IN2: Process value ch. 2
MATH: Process value maths.
IN 1
* Total decimal point MADP
Selection of the mathematics
channel decimal point
Selection range:
0 to 4 decimal points
9999,9
* Total offset MAOFF
Signal offset/ measured
value displacement of the
mathematics channel
Value:
-19999 to 99999
0,0
* X-value for linearisation
point 1
MSCLO
Input value (X) for the first
point in the linearisation table
Value:
-19999 to 99999
0,0
* X-value for the last
linearisation point
MSCHI
Input value (X) for the last
point in the linearisation table
Value:
-19999 to 99999
100,0
* Decimal point Y-value
MDIDP
Selection of the decimal point
for the Y table value
Selection range:
0 to 4 decimal points
9999,9
* Y-value linearisation
point 1
MDILO
Output value (Y) for the first
point on the linearisation table
Value:
-19999 to 99999
0,0
* Y-value last linearisation
point
MDIHI
Output value (Y) for the last
point on the linearisation table
Value:
-19999 to 99999
100,0
* Table offset
MDIOF
Measured value offset on the
linearisation values
Value:
-19999 to 99999
0,0
Process transmitter
53