Installation
Hardware Installation and Maintenance Manual 3
13
4. Strip back the jacket and insulation of the current loop and serial
cable just enough to connect to the mating terminal block, shown in
Figure 3–6 below. The mating terminal block can be pulled up and
removed from its base to make the cable connection process easier.
5. Connect the 4-20 mA current loop signal wires to the appropriate
terminals, as indicated in Table 3–1 on page 3–14.
6. Connect the serial cable wires to the appropriate terminals according
to Table 3–1 on page 3–14. For reference, Table 3–1 on page 3–14
also shows the corresponding pin numbers for configuring a nine-pin
Sub-D connector for connection to a computer serial port.
7. Re-insert the mating terminal block into its base and verify that each
connection is secure.
8. Strip back the jacket and insulation of the alarm cables just enough
to connect to the alarm relays, shown in Figure 1–6 on page 1–11 or
Figure 1–7 on page 1–12.
9. Connect the cable for the General Fault Alarm to the common and
NO terminals on the corresponding relays shown in Figure 1–6 on
page 1–11 or Figure 1–7 on page 1–12.
10. Connect the cable for the Concentra High Alarm to the common
and NO or NC terminals (for OPEN or CLOSED, respectively, when
activated) on the corresponding relay shown in Figure 1–6 on page
1–11 or Figure 1–7 on page 1–12.
11. Close and tighten the Analyzer A electronics enclosure cover.
The external serial cable included in the shipping container is
provided for service and troubleshooting purposes and is not
intended for permanent installation.
Figure 3–6 Analyzer A mating terminal
block (TB2) for connecting signal cables
TB2
121110987654321

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser SpectraSensors 2-Pack Analyzer System and is the answer not in the manual?

Endress+Hauser SpectraSensors 2-Pack Analyzer System Specifications

General IconGeneral
BrandEndress+Hauser
ModelSpectraSensors 2-Pack Analyzer System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals