29
Measurement deviations according to Eppendorf AG
Eppendorf Reference
®
2
English (EN)
8 Measurement deviations according to Eppendorf AG
8.1 Single-channel pipette with fixed volume
Model Testing tip
epT.I.P.S.
Error of measurement
Systematic error Random error
± % ± μL ± % ± μL
1 µL 0.1 µL – 10 µL
dark gray
34 mm
2.5 0.025 1.8 0.018
2 µL 2.0 0.04 1.2 0.024
5 µL 0.1 µL – 20 µL
medium gray
40 mm
1.2 0.06 0.6 0.03
10 µL 1.0 0.1 0.5 0.05
20 µL 0.5 µL – 20 µL L
light gray
46 mm
0.8 0.16 0.3 0.06
10 µL 2 µL – 200 µL
yellow
53 mm
1.2 0.12 0.6 0.06
20 µL 1.0 0.2 0.3 0.06
25 µL 1.0 0.25 0.3 0.075
50 µL 0.7 0.35 0.3 0.15
100 µL 0.6 0.6 0.2 0.2
200 µL 0.6 1.2 0.2 0.4
200 µL 50 µL – 1000 µL
blue
71 mm
0.6 1.2 0.2 0.4
250 µL 0.6 1.5 0.2 0.5
500 µL 0.6 3.0 0.2 1.0
1000 µL 0.6 6.0 0.2 2.0
2.0 mL 0.5 mL – 2.5 mL
red
115 mm
0.6 12 0.2 4
2.5 mL 0.6 15 0.2 5