8.3 Multi-channel pipette
8.4 Test conditions
Test conditions and test analysis in accordance with the standard ISO 8655, Part 6. The
inspections are performed with a calibrated analytical balance with evaporation
protection.
Number of determinations per volume: 10
Water according to ISO 3696
Inspection at 20 °C – 27 °C ±0.5 °C
Dispensing onto the tube inner wall
0.5 mL – 5 mL 0.1 mL – 5 mL L
violet
175 mm
0.5 mL 5.0 25 1.0 5.0
2.5 mL 3.0 75 0.9 22.5
5.0 mL 2.0 100 0.8 40
1 mL – 10 mL 1 mL – 10 mL
turquoise
165 mm
1.0 mL 3.0 30 0.6 6
5.0 mL 0.8 40 0.2 10
10.0 mL 0.6 60 0.15 15
1 mL – 10 mL 1 mL – 10 mL L
turquoise
243 mm
1.0 mL 6.0 6 1.0 10
5.0 mL 3.0 150 0.9 45
10.0 mL 2.0 200 0.7 70
Model Testing tip
epT.I.P.S.
Testing
volume
Error of measurement
Systematic error Random error
± % ± μL ± % ± μL
0.5 µL – 10 µL 0.1 µL – 20 µL
medium gray
40 mm
0.5 µL 12.0 0.06 8.0 0.04
1 µL 8.0 0.08 5.0 0.05
5 µL 4.0 0.2 2.0 0.1
10 µL 2.0 0.2 1.0 0.1
10 µL – 100 µL 2 µL – 200 µL
yellow
53 mm
10 µL 3.0 0.3 2.0 0.2
50 µL 1.0 0.5 0.8 0.4
100 µL 0.8 0.8 0.3 0.3
30 µL – 300 µL 20 µL – 300 µL
orange
55 mm
30 µL 3.0 0.9 1.0 0.3
150 µL 1.0 1.5 0.5 0.75
300 µL 0.6 1.8 0.3 0.9
Model Testing tip
epT.I.P.S.
Testing
volume
Error of measurement
Systematic error Random error
± % ± μL ± % ± μL