EasyManua.ls Logo

eschmann T20 Series - Post-Procedure Care and Cleaning; After-Use Procedures; Cleaning Operation Table and Accessories; Cleaning Mattresses and Pads

eschmann T20 Series
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
P44/62 T-IM114b
8.0 AFTER USE, CLEANING & CARE
8.1 After use procedures
After each procedure the table should be cleaned and
disinfected in accordance with local procedures which take
into account the guidelines provided in section 8.2 and 8.3
of this manual.
Accessories should be removed from the table and after
cleaning and disinfection the head and leg section should
be fully lowered as good practise.
The table battery LEDs should be checked to see if the
batteries require recharging, recharge if applicable. If the
preceding procedure was the last one for the day or shift,
then the batteries should be placed on charge as a routine
procedure (see section 5.3.1.2)
The table must be switched ‘off’ ( ) at the table on/off switch
(item 5 of Fig. 2.2) when the table is not in use. The green
LED item 4 Fig. 2.2 will not be illuminated to show the
table is ‘off’.
If the table has been left switched ‘on’ and no movement
command has been given via any controller (hand, foot or
standby) for a period of six hours, an audible warning sounds
(a slow repeating double ‘beep’). This is provided to warn
the user to switch the table ‘off’ ( ) at the table on/off switch
(item 5 of Fig. 2.2). If the table is still in use when this alarm
sounds (e.g. table being used for a long procedure) and the
user does not want the table switched ‘off’
(e.g. Trendelenburg may be required urgently) the alarm
can be stopped by pressing any control input briefly, or,
switching the table ‘off’, waiting 10 seconds and then
switching the table back ‘on’. The alarm will sound again in
six hours if no movement command has been given and
the table has not been switched ‘off’. This feature can be
disabled or the time period of six hours can be altered,
contact Eschmann Equipment for more information.
8.2 Cleaning
NOTE
Eschmann Equipment cannot accept liability
for the efficiency or the effects of any cleaning
or disinfection techniques. Only use the
materials detailed in the following sections.
WARNING
Disconnect from the mains electrical supply
before cleaning or disinfecting the operation
table, mattresses, or accessories. When
cleaning with a brush wear suitable eye
protection (brushes are prone to ‘flick’
particles) and at all times wear suitable
personal protection (e.g. gloves and overalls).
CAUTION
Do not immerse the handsets or footswitch in
liquids when cleaning or disinfecting.
If you have any particular cleaning problems, contact the
Eschmann Equipment After Sales Service Department at
the address given inside the front cover of this Manual.
8.2.1 Operation table and accessories
Fully raise the tabletop, remove all accessories and all the
mattresses, then wash the table with hot (55°C) neutral
(pH7) detergent solution (diluted in accordance with the
manufacturers instructions) and rinse with clean water. Use
a small brush to clean areas of difficult access.
Clean non-electrical accessories as detailed above for the
table but only use a damp cloth to clean the handsets and
footswitch.
8.2.2 Mattresses and pads
WARNING
Mattresses and pads should be checked for any
cuts or tears in the outer covering. If any of
these defects are found, the mattress must be
replaced to eliminate possible biological
hazards.
CAUTION
Remove spilt ether and anaesthetic liquid from
mattresses and pads immediately.
When cleaning mattresses and pads:
v DO NOT use phenolic disinfectants.
v DO NOT use abrasive cleaning agents.
v DO NOT use organic solvents (e.g. petroleum spirit,
carbon tetrachloride, or tetra-chloromethane).
v DO NOT allow petroleum-based oil, vegetable oil,
wax, or grease to remain on the mattress.
v DO NOT dry mattresses with direct heat (e.g.
radiators, electric fires, or hot-air blowers) allow them
to dry at normal room temperatures.
Wash mattresses and pads thoroughly, with hot (55°C)
neutral (pH7) detergent solution (diluted in accordance with
the manufacturers instructions) and then rinse with clean
water. Use a small brush to clean areas of limited access.
Remove stubborn stains and deposits with good quality
vinyl cleaner using a soft-bristle brush. Dry all surfaces
with absorbent paper, to avoid damage do not leave them
wet in contact with another surface (e.g. the table top).

Table of Contents

Related product manuals