T-IM114b P45/62
T20
Series
OPERATION TABLE
8.3 Disinfection
NOTE
All equipment and accessories returned to
Eschmann Equipment must be accompanied
by a Decontamination Certificate, signed by an
authorized person of managerial status.
Appendix 2 shows an example of a suitable
Decontamination Certificate that can be
photocopied for use.
CAUTION
Hypochlorite solutions can damage metal
parts, after disinfection, ensure it is rinsed of
thoroughly with water. Prolonged exposure to
hypochlorite solutions may degrade mattress
material. Avoid hypochlorite solutions and
other liquids coming into contact with any
internal parts of the table.
8.3.1 Disinfection procedure
A well ventilated area should be designated and used as the
disinfection area. Access to the area should be restricted to
those people involved in the disinfection process.
The following disinfection procedure is used by Eschmann
Equipment, and its use is recommended if no other local
approved procedures are available.
8.3.2 Table, accessories mattresses and pads
Disinfect the operation table, accessories, mattresses and
pads as follows:
i Disassemble the table and/or accessories as far
as possible without the use of tools, remove
mattresses and pads.
ii Scrub all surfaces and crevices with hot (55°C)
neutral (pH7) detergent solution (diluted in
accordance with the manufacturers instructions) to
remove all visible contamination. Use a small brush
to clean areas of limited access repositioning the
table as required to gain access to all surfaces. For
stubborn mattress stains see section 8.2.2.
iii Wash down with hot (55°C) water.
iv Dry all surfaces with absorbent paper.
v Wash down all surfaces and crevices with one of
the solutions below:
v a 70% solution of industrial methylated spirit in
water.
v a 1000 to 5000mg/litre solution of hypochlorite
in water (see Caution note above).
vi Wash down thoroughly with clean water.
vi Dry all surfaces with absorbent paper, to avoid
damage do not leave mattresses wet in contact with
another surface.
vii Dispose of all cleaning material and solutions in
accordance with authorized disposal procedures.
8.4 Care
WARNING
The head and leg section gas-springs are filled
with gas at high pressure, do not try to open
them. The gas-springs should be replaced
immediately any signs of leaking or
deterioration in performance are noted (e.g.
movement of section when locked).
CAUTION
Do not lubricate the head or the
leg section, gas-springs.
Once a week:
i Remove fluff and debris from the head and leg
section guide pin location sockets and if required
spray the pins and into the sockets with a suitable
light lubricant such as WD40.
ii Remove all mattresses and check them for any cuts,
scuffs or other damage and replace as required.
iii Check that the table covers are not cracked,
chipped, or otherwise damaged and arrange
replacement as required.
iv Check the table for any signs of wear or damage
that requires attention and arrange remedial action
if required.
v Ensure batteries have been regularly charged and
charge them if required.
vi On the T20-m table clean the castors and remove
any debris caught between them.
Once every six months the table should receive a safety
check and service as detailed in the service manual (which
provides full details of part replacement, safety checks and
routine maintenance). During this service the tables
calibration should be checked and if required it should be
recalibrated. Also at least once every year the electrical
conductivity of the table should be checked. To arrange
these contact the Eschmann After Sales Service
Department, see inside front cover for contact details.