EasyManua.ls Logo

EST Centurion WS - Page 74

Default Icon
205 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure 4ii: Wash Station Calibration Screen
10. The Left/Right (L/R), Front/Rear (F/R) or Up/Down arrows are to the left of the
screen. One click of the L/R or F/R arrow will move the needle 20 steps in the
corresponding direction. While clicking the Down arrow will move 100 steps and the
Up arrow will move 20 steps. (Note: 20 steps = 0.5mm). In order to move the
needle axis into the calibration position <click> on the double arrow.
11. When the needle is calibrated in the wash station, <click> Update Calibration. See
Figure 4ii.
12. When Update Calibration is clicked the software will automatically proceed to the
next step in the calibration routine.
13. The next step is Vial 1 position calibration. The needle will move to vial position 1
and the software will prompt you with a pop-up, see Figure 4jj.
Figure 4jj: Vial 1 Calibration Pop-up
14. <click> OK and the needle position over Vial 1 can be calibrated.
15. The center point of Vial 1 is based off of the center point of the wash station
calibration. If this needs to be adjusted, the position arrows can be adjusted as they

Related product manuals