EasyManuals Logo

FAST ComTec MCDWIN User Manual

Default Icon
59 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
Windows Server Program
ComTec GmbH 4-6
cmline7=3
cmline8=Ge
cmline9=test
range=2048 ; subset of parameters as in a WEBMCAA.CFG file...
rtpreset=1000
rtprena=0
ltpreset=1000
ltprena=0
roipreset=0
roiprena=0
roimin=0
roimax=2048
autoinc=0
datname=C:\MCA\DATA\SD0002.mcd
savedata=1
fmt=dat
caluse=1 ; Use Energy calibration
calch00=1172.00 ; Calibration points..
calvl00=1173.264000
calch01=1331.00
calvl01=1332.500000
caloff=-0.506315 ; Calibration formula polynomial coefficients..
calfact=1.000750
calfact2=0
calfact3=0
calunit=keV ; Calibration unit
roi=1156 1199 ; ROI limits..
peak=1173.230000 ; calibration peak in this ROI
roi=1323 1353
peak=1332.480000 :
The following commands perform actions and therefore usually are not included in a
WEBMCAx.CFG or .MCD file:
start ; Clears the data and starts a new acquisition for system 1. Further
; execution of the .CTL file is suspended until any acquisition stops
; due to a preset.
halt ; Stops acquisition of system 1 if one is running.
cont ; Continues acquisition of system 1. If a time preset is already
; reached, the time preset is prolongated by the value which was valid
; when the „start“ command was executed. Further execution of the ;.CTL
file is suspended (see start).
savecnf ; Writes the settings into MCAA.CFG

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the FAST ComTec MCDWIN and is the answer not in the manual?

FAST ComTec MCDWIN Specifications

General IconGeneral
BrandFAST ComTec
ModelMCDWIN
CategoryComputer Hardware
LanguageEnglish