14 Data Logger and Transmission of Data FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 105
RS232
default: 9600 bits/s, 8 data bits, even parity, 2 stop bits, protocol RTS/CTS (hardware,
handshake)
The transmission parameters of the RS232 interface can be changed:
14.2.5 Online Transmission of Data to a Terminal Program
Start the terminal program.
Enter the transmission parameters into the terminal program (see section 14.2.4). The
transmission parameters of the terminal program and of the transmitter have to be
identical.
Select the program branch Output Options. Press ENTER.
Select the channel for which the online transmission of data is to be activated. Press
ENTER until the menu item Serial Output is displayed.
Set the storage rate (see section 14.1.2).
Start the measurement. The measuring point number will be requested (see section
14.1.4).
14.2.6 Offline Transmission of Data to a Terminal Program
Start the terminal program.
Enter the transmission parameters into the terminal program (see section 14.2.4). The
transmission parameters of the terminal program and of the transmitter have to be
identical.
Enter HotCode 232-0- immediately after the transmitter has been switched on.
Set the transmission parameters in the 4 scroll lists. Press
EN-TER.
baud: baud rate
data: number of data bits
par: parity
st: number of stop bits
Select yes to activate the online transmission of data.
The measured values are transmitted during the
measurement.
Select Special Funct.\Print Meas.Val.. Press
ENTER.
baud<data par st
9600 8bit EVEN 2
Serial Output
no >YES<
SEND ONLINE-HEAD
20 mm
Special Funct.
Print Meas.Val.