11 Basic Measurement FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 77
• Select a channel with key and .
• Press key to activate or deactivate the selected channel. Press ENTER.
A deactivated channel will be ignored during the measurement. Its parameters will re-
main unchanged.
If the data logger or the serial interface is activated, the measuring point number must be
entered:
11.5 Defining the Number of Sound Paths
The number of transits of the ultrasonic waves through the medium depends on the
placement of the transducers on the pipe.
If the number of transits is odd (diagonal mode), the transducers will be mounted on op-
posite sides of the pipe.
If the number of transits is even (reflection mode), the transducers will be mounted on the
same side of the pipe.
The channels for the measurement can be activated and
deactivated.
: the channel is active
–: the channel is not active
•: the channel can not be activated
This display will not be indicated if the transmitter has only
one measuring channel.
Note! A channel can not be activated if the parameters are not valid, e.g. if
the parameters in the program branch Parameter of the channel
are not complete.
Enter the measuring point number. Press ENTER.
If arrows are displayed in the lower line on the right, ASCII
text can be entered. If no arrows are displayed, only digits,
point and hyphen can be entered.
A:Meas.Point No.:
xxx (↑↓← →)