14 Data Logger and Transmission of Data FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 109
14.2.8 Structure of the Data
The header is transmitted at the beginning of the measurement. The first 4 lines contain
general information about the transmitter and the measurement. The following lines con-
tain the parameters of each channel.
The line \DATA will be transmitted next. Afterwards the column titles will be transmitted
for the respective channel (see Tab. 14.3) The measured values are transmitted after-
wards.
In every storage interval, one data line per activated measuring channel is transmitted.
The line "???" will be transmitted if there are no measured values available for the sto-
rage interval.
Example: \DEVICE :
G60X-XXXXXXXX
\MODE : ONLINE
DATE : 2011-01-09
TIME : 19:56:52
Par.Record
Meas.Point No.: : A:F5050
Pipe
Outer Diameter : 60.3 mm
Wall Thickness : 5.5 mm
Roughness : 0.1 mm
Pipe Material : Carbon Steel
Lining : WITHOUT LINING
Medium : Natural gas user
Medium Temperat. : 38 C
Fluid pressure : 60.00 bar
Transducer Type : xxx
Sound Path : 3 NUM
Transd. Distance : -15.6 mm
Damping : 20 s
Full-Scale Val. : 4.50 m3/h
Physic. Quant. : act. Volume flow
Unit Of Measure : [m3/h]/[m3]
Numb.Of Meas.Val : 100
Example: \DATA
A: \*MEASURE; Q_POS; Q_NEG;
B: \*MEASURE; Q_POS; Q_NEG;
example: With a storage interval of 1 s, 10 lines "???" will be transmitted if the
measurement has been restarted after a 10 s interruption for the positio-
ning of the transducers.