B Technical Data FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 193
explosion protection
A
T
E
X
category
EPL
zone
gas: 3G dust: 2D
Gc Db
2 21
marking
0637
II3G Ex nA nC ic IIC (T6)T4 Gc
II2D Ex tb IIIC T 100 °C Db
T
a
-10
...+(
50)60 °C
certification IBExU10ATEX1067
type of protection gas: non sparking
dust: protection by enclosure
temperature inputs: intrinsic safety
measuring functions
physical quantities operating volumetric flow rate, standard volumetric flow rate, mass flow
rate, flow velocity
totalizer volume, mass
calculation functions average, difference, sum
diagnostic functions sound speed, signal amplitude, SNR, SCNR, standard deviation of ampli-
tudes and transit times
data logger
loggable values all physical quantities, totalized values and diagnostic values
capacity > 100 000 measured values
communication
interface RS232/USB
serial data kit
software (all Win-
dows™ versions)
- FluxData: download of measurement data, graphical presentation,
conversion to other formats (e.g. for Excel™)
- FluxKoef: creating medium data sets
сable RS232
adapter RS232 - USB
transport case
dimensions 500 x 400 x 190 mm
outputs
The outputs are galvanically isolated from the transmitter.
accessories output adapter (optional)
current output
range 0/4...20 mA
accuracy 0.1 % of reading ±15 μA
passive output U
ext
= 4...9 V, depending on R
ext
R
ext
< 200 Ω
binary output
optorelay 26 V/100 mA
binary output as alarm
output
- functions limit, change of flow direction or error
binary output as pulse
output
- pulse value 0.01...1 000 units
- pulse width 1...1 000 ms
FLUXUS G608**-A2