B Technical Data FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 217
Lamb Wave Transducers (zone 2)
technical type GRM1NH3 GRP1NH3 GRQ1NH3
order code GLM-NA2NL GLP-NA2NL GLQ-NA2NL
transducer frequency MHz 1 2 4
medium pressure
1
min. extended bar - - -
min. bar metal pipe:
10 (d > 60 mm)
5 (d < 60 mm)
plastic pipe: 1
metal pipe:
10 (d > 35 mm)
5 (d < 35 mm)
plastic pipe: 1
metal pipe:
10 (d > 15 mm)
5 (d < 15 mm)
plastic pipe: 1
inner pipe diameter d
min. extended mm 30 15 7
min. recommended mm 40 20 10
max. recommended mm 90 50 22
max. extended mm 150 70 35
pipe wall thickness
min. mm 2 1 0.5
max. mm 5 3 1
material
housing PPSU 
with stainless 
steel cap and 
transducer shoe
 304 
(1.4301)
PPSU 
with stainless 
steel cap and 
transducer shoe
 304 
(1.4301)
PPSU 
with stainless 
steel cap and 
transducer shoe
 304 
(1.4301)
contact surface PPSU PPSU PPSU
degree of protection 
according to IEC/
EN 60529
IP65 IP65 IP65
transducer cable
type 1699 1699 1699
length m 4 4 3
dimensions
length l mm 84 84 70
width bmm404030
height h mm 59 59 47.5
dimensional drawing
operating temperature
min. °C -40 -40 -40
max. °C +170 +170 +170
temperature
compensation
xxx