EasyManua.ls Logo

Flexim FLUXUS G800 - Page 102

Flexim FLUXUS G800
161 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FLUXUS G80x 14 Data Logger and Transmission of Data
102 UMFLUXUS_G8V4-6-1EN, 2018-10-10
14.2.5 Online Transmission of Data to a Terminal Program
Start the terminal program.
Enter the transmission parameters into the terminal program (see section 14.2.4). The transmission parameters of the
terminal program and the transmitter have to be identical.
Settings in the Transmitter
Select the program branch Output Options. Press ENTER.
Select the channel for which the online transmission of data is to be activated. Press ENTER until the menu item Serial
Output is displayed.
Set the storage rate (see section 14.1.2).
Start the measurement. The measuring point number will be requested (see section 14.1.4).
14.2.6 Offline Transmission of Data to a Terminal Program
Start the terminal program.
Enter the transmission parameters into the terminal program (see section 14.2.4). The transmission parameters of the
terminal program and the transmitter have to be identical.
Select yes to activate the online transmission of data. Press ENTER.
The measured values are transmitted during the measurement.
Select Special Funct.\Print Meas.Val.. Press ENTER.
This error message will be displayed if no measured values are stored.
Press ENTER.
This message will be displayed if the measurement values are transmitted.
The progress of the transmission of data is displayed by a bar graph.
This error message will be displayed if an error has oc-curred during the serial transmission.
Press ENTER. Check the connections and make sure that the PC is ready to receive data.
Serial Output
no >YES<
SEND ONLINE-HEAD
20 mm
Special Funct.
Print Meas.Val.
NO VALUES !
Print Meas.Val.
Send Header 01
................
■■■■■■
................
SERIAL ERROR !
Print Meas.Val.

Table of Contents

Related product manuals