FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019 Page 63 of 170
[ 5-- SCALE ] Scale Block
This block is related to the measurement related parameters, which describe the use of the scale. The most
important is parameter 511, which limits the use of the parameters in main blocks 5, 6 and 8. The selections
of this parameter are;
Selections of parameters at set up and calibration are free
for industrial usage of the instrument.
Industrial weighing
only gross
Auto zeroing, taring and power on zero are disabled.
(recommended for tank /silo weighing in gross.)
Industrial weighing net
(taring can be enabled)
Auto zeroing and power on zero are disabled.
(recommended for tank /silo weighing in gross.
Approved scale
according to OIML
Metrology related parameters are restricted to limits of OIML
R76 and EU type approval of the instrument.
After setting parameter 511, even you set any parameter out of the accepted range of selection of par 511, it
is saved in the acceptable limit. For example, if taring is activated at INDG selection, it will be disabled while
exiting from set up. For approved scales, set the parameters in main blocks 5, 6 and 8 perform calibration
carefully due to sealing of the scale in legal usage.
SCALE RELATED PARAMETERS MAIN BLOCK
Press key sequentially to access this main block,
or press key or key to enter configuration parameters,
or press key to enter the next block,
or press key to exit from programming.
SCALE SET UP
Press key or key again to enter this menu.
Or press key to go to the next sub-block.
Industrial. All parameters can be freely selected.
Industrial weighing of tank, hopper or silo in gross.
(Taring, AZTrack and Power on Zero are disabled)
Industrial weighing in Net of tank, hopper or silo.
(disabled AZT and Power on Zero, enabled Tare)
OIML approved scale.
Temporary indication with key.
Toggle. Start and end high resolution by pressing
key in sequence.
Always high resolution
Disable.
± 2%
± 2%, [POWER ON ZERO ERR] prompt cannot be
erased. Call service.
± %10