Table 7-1. Error Messages (cont.)
Error Code Error Message Causes and Solutions
Error 422
Not a totalizer
channel
Cause: Error occurs when a user attempts to set a TOT
channel on any channel other than Ch402.
Solution: Check the parameter value and try again.
Error 423
Not an analog
channel
Cause: Error occurs when a user attempts to set an analog
channel on any channel other than Chx01 through Chx22.
Solution: Check the parameter value and try again
Error 424
3-wire or 4-wire not
allowed on channel
Cause: Channel cannot be configured for 3-wire or 4-wire
because (1.) the input type does not support it (2.) the
channel is not a dedicated sense channel.
Solution: Check to make sure that the sense wires are in the
sense terminals and not the input terminals. When a sense is
configured, it is configured on the channel associated with the
input terminal and automatically reserves the sense channel
directly across from the input. See Chapter 2 and Chapter 3
for more information.
Error 425
Function not allowed
on current channel
Cause: Chx21 or Chx22 can only be configured as a current
channel.
Solution: Set up the input on a different channel.
Error 427
Invalid current
channel
Cause: An Input Module was swapped with an external Input
Module while a scan was paused.
Solution: Reinstall the Input Module that was used when the
scan was started.
Error 428
Parameter value out
of range
Cause: Parameter in command exceeds the limit.
Solution: Check the parameter value and try again.
Error 429
Channel limit
exceeded
Cause: A single math channel is configured to use more than
10 channels.
Solution: Limit the number of math channels to no more than
10 with the CALCulate:MATH:SOURce command.
Error 430 String too long
Cause: Too many characters are included in a string
parameter. For example,Probe ID (max. 24 characters); Unit
(max. 3 characters); Channel label (max. 16 characters); Test
setup file name (max. 8 characters).
Solution: Check the parameter value and try again.