Fluke 1586A - Measurement Specifications

Fluke 1586A
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Product Overview and Specifications
Measurement Specifications 1
1-15
Measurement Specifications
Accuracy specifications generally apply with medium and slow sample rates (unless otherwise noted), after a warm-up
time of 1 hour, and within an environment temperature range of 18 °C to 28 °C, and may depend on the channel. The
confidence level for accuracy specifications is 95 % within 1 year of calibration.
Scan rate
Fast ..................................................................... 10 channels per second max (0.1 seconds per channel)
Medium ............................................................... 1 channel per second (1 second per channel)
Slow .................................................................... 4 seconds per channel
Display Resolution ................................................ 4 ½ to 6 ½ digits, depending on function and Sample Rate (see
Measurement Characteristics tables below to find the display
resolution of temperature readings)
PRT/RTD
Temperature Range .............................................. 200 °C to 1200 °C (depending on the sensor)
Resistance Range ................................................. 0 Ω to 4 kΩ
Offset Compensation
0 Ω to 400 Ω, 4-wire........................................ automatic current reversal
400 Ω to 4000 Ω or 3-wire .............................. none
Source Current Reversal Interval (0 Ω to 400 Ω range)
Fast sample rate ............................................. 2 ms
Medium sample rate ....................................... 250 ms
Slow sample rate ............................................ 250 ms
Maximum Lead Resistance (4-wire Ω) ................ 2.5 % of range per lead for 400 Ω and 4 kΩ ranges.
PRT/RTD Resistance Accuracy
Accuracy is given as ± (% of measurement + ohms). Basic accuracy is for 4-wire PRT/RTD using medium or slow sample
rate. When using 3-wire PRT/RTD add 0.013 Ω to the accuracy specification for internal resistance mismatch and voltage
offset if using Channel 1, or add 0.05 Ω if using channels x01 through x20. When using Fast sample rate add the number
given in the table to the accuracy specification. The 4 kΩ range accuracy specification can be decreased by 0.06 Ω if
using Slow sample rate. If the environment temperature is outside the specified range, multiply the temperature coefficient
numbers by the temperature deviation and add to the accuracy specification.
Range DAQ-STAQ Module High-Capacity Module Fast Sample Rate
T.C./ °C Outside 18 °C to
28 °C
0 Ω to 400 Ω 0.002 % + 0.002 Ω 0.003 % + 0.003 Ω add 0.003 Ω 0.0001 % + 0.0008 Ω
400 Ω to 4 kΩ 0.004 % + 0.1 Ω
0.006 % + 0.1 Ω
add 0.08 Ω 0.0001 % + 0.008 Ω