Product Overview and Specifications
How to Contact Fluke Calibration 1
1-13
How to Contact Fluke Calibration
To contact Fluke Calibration, call one of the following telephone numbers:
Technical Support USA: 1-877-355-3225
Calibration/Repair USA: 1-877-355-3225
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31-40-2675-200
Japan: +81-3-6714-3114
Singapore: +65-6799-5566
China: +86-400-810-3435
Brazil: +55-11-3759-7600
Anywhere in the world: +1-425-446-6110
To see product information and download the latest manual supplements, visit Fluke
Calibration’s website at www.flukecal.com.
To register your product, visit http://flukecal.com/register-product.
Calibration and Repair Information
To schedule and send the Product to Fluke for calibration or repair:
1. Contact the Fluke Service Center in your area to schedule the calibration or repair
(see “Contact Fluke Calibration”).
2. Pack and secure the Product in a shipment box with a minimum of 2 inches of
packing around the Product to prevent damage.
3. Send the Product to the Service Center.
General Specifications
Mains Voltage
100 V Setting ...................................................... 90 V to 110 V
120 V Setting ...................................................... 108 V to 132 V
220 V Setting ...................................................... 198 V to 242 V
240 V Setting ...................................................... 216 V to 264 V
Frequency .............................................................. 47 Hz to 440 Hz
Power Consumption ............................................. 36 VA peak (24 W average)
Environment Temperature
Operating ............................................................ 0 °C to 50 °C
Full accuracy ....................................................... 18 °C to 28 °C
Storage ............................................................... 20 °C to 70 °C
Warm-up ............................................................. 1 hour to full accuracy specifications
Relative Humidity (non-condensing)
Operating ............................................................ 0 °C to 30 °C <80 %
30 °C to 50 °C <50 %
Storage ............................................................... 20 °C to 70 °C <95 %