3-1
Chapter 3
Input and Channel Configuration
Introduction
This chapter supplies instructions on how to wire inputs to the Input Module then
configure the associated channel.
Input Wiring
The Universal Input Module
The 2686A-100 Universal Input Module (the Input Module) is used to wire inputs of
various types to the Product (see Figure 3-1). Each Input Module has 22 analog channels
(ChX01 through ChX20) that can be configured to measure dc volts, ac volts, resistance,
frequency, and temperature and two low-burden current channels (ChX21 and ChX22)
that can be configured to measure ac and dc current without the use of an external shunt
resistor.
Note
To expand the functionality of the Product, current sources can also be connected
to channels ChX01 through ChX20 by the use of an accessory shunt resistor. See
Chapter 6 for part number and ordering information.
1.888.475.5235info@Fluke-Direct.com
Fluke-Direct.com

Table of Contents

Other manuals for Fluke 2638A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 2638A and is the answer not in the manual?

Fluke 2638A Specifications

General IconGeneral
Resistance Range0 to 100 MΩ
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz
Voltage Range±300 V
Resolution6.5 digits
MemoryUSB
InterfaceUSB, Ethernet
DisplayColor
Measurement TypesDC voltage, AC voltage, resistance, DC current, AC current, frequency, temperature