Operating Features
5
Understanding the Pushbuttons
Use the pushbuttons to control operation of the tester.
Button Function
BOND
200mA
Start the 200 mA Earth Bond test.
BOND
25A
Start the 25 A Earth Bond test.
INSUL-
ATION
Start the Insulation test.
I
SUB
Start the Substitute Leakage Current test.
LOAD/
LEAK
Start the combined Load/Earth Leakage
Current test.
I
TOUCH
Start the Touch Current test.
STOP
Stop the current action and return to Idle
screen.
IEC
LEAD
Start the IEC Lead test.
ZERO
Start zeroing the Earth Bond test.
MEM
Store test results.
CLEAR
Clear stored data.
PRINT
Print test results.
Understanding the Beeper Sounds
The tester can make several types of beeper sounds.
Sound Meaning
Click A button is pressed.
1 beep A test passed.
2 beeps close
together
- A test failed.
- Warning, see display.
- The STOP button is pressed,
the current action is aborted.
1 long beep A continuous non-live test has
been started.
2 beeps + 1 long
beep
A continuous live test has been
started.
Continuous 2
tone sound
Dangerous condition ! Unplug the
unit immediately!