DISPLAY EXPLANATION
+ number
Dangerous tester failure. Unplug the
tester, prevent it from being used, and
contact Fluke for repair.
U
N-PE
The Neutral-Earth voltage is
dangerously high. Unplug the tester!
The mains polarity is incorrect. Unplug
the tester!
U
N-PE
Mains supply earth connection is
missing/open circuit. Unplug the
tester!
Setting Up the Tester
The only requirement to set up the tester, is to zero the
earth bond test.
Zeroing the Earth Bond Test
For correct earth bond test results you must zero the
earth bond lead to eliminate its resistance:
• when setting up your new tester. Earth bond tests are
locked out unless the bond zero icon ∅ is on.
• occasionally, dependent on the condition of the bond
socket and the test lead plug a dirty plug/socket can
result in a significant contact resistance.
To zero the test lead, do the following:
1
Attach the crocodile clip to the test lead
and insert the test lead plug into the
BOND 25A/200mA socket, see Figure 3.
2
Firmly attach the crocodile clip to the
BOND ZERO bar on the tester.
3
ZERO
Press zero. The display shows the test
progress, see the table on the next page.