Maintaining the Tester
29
Maintaining the Tester
Cleaning
Periodically wipe the case with a damp cloth and mild
detergent. Do not use abrasives or solvents.
Dirt or moisture on the earth bond test lead plug can
result in a contact resistance that affects the readings.
Therefore periodically zero the earth bond test (see page
10).
Calibration
To ensure the accuracy of the tester is maintained at high
level it is recommended that the tester is calibrated at
least once every 12 months. Calibration must be carried
out by qualified personnel. Contact your local Fluke
representative for calibration ( see Contacting Fluke on
page 1).
Accessories
Table 2 and Table 3 list the part numbers of the
accessories.
To order the accessories contact your local Fluke
representative (see Contacting Fluke on page 1).
Table 2. Standard Accessories
Item Part Number
Alligator Clip 532269474055
Test Lead 532269474056
Test probe for touch current 1276841
Users Manual (this manual)
1)
1)
Can be downloaded from your regional Fluke website,
start at www.fluke.com .
Table 3. Optional Accessories
Item Part
Number
SP1000 Mini Printer 1597281
EXTL100 Extension Lead Test Adapter 2414348
TA700 Appliance Adapter for 110V Tools 2389678
Fluke PowerPAT Plus Appliance Testing
Software
2143155
Note:
The PowerPat plus software cannot support
direct data input via the RS-232 interface. The
test results must be entered manually.