Fluke 6200 - IEC Lead Test Procedure

Fluke 6200
39 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6200
Users Manual
24
IEC Lead Test
The IEC lead test tests the IEC lead for:
Earth bond resistance
Insulation resistance Live-Neutral against earth.
Live-Neutral lead/fuse continuity and polarity.
If there is a swapped polarity condition and a continuity
failure in the same test, a failed polarity message will be
displayed.
You can use the adapter EXTL100 (optional accessory)
to test extension leads.
To perform the IEC Lead test, do the following:
1
Connect the IEC lead as indicated on the
tester, see also Figure 9.
2
IEC
LEAD
Start the test.
The IEC Lead test runs only in the single
test mode.
The display shows the test progress, see the
table on the next page.
3
MEM
Store the test result, if required.
Figure 9. IEC-Lead Test Connections