Users Manual
Testing Appliances
23
LOAD A
> A
VA
> kVA
>
The load current is 0.2 A.
Load current overrange detected.
The load power is 50 VA.
Load power overrange detected.
Excess load surge condition detected.
PE LEAK mA
> mA
>
The leakage current is 0.2 mA
Leakage current overrange.
Excess leakage current surge condition
detected.
Limit
The acceptable test limit may have been
exceeded. Refer to standards and/or
guidance material.
FAIL X
The Fuse/L-N Loop pre-test tailed.
Check that the appliance power switch
is on. See also Fuse/L-N Loop Pre-test
below.
Fuse/L-N Loop Pre-test
The pre-test verifies the fuse and lead continuity by
applying a low voltage signal across the appliances
phase and neutral pins.
Very low power appliances, or appliances with
electronically controlled on/off switches or with an
inductance may fail this test. In this case the symbol
FAIL X will appear.
To enable you to test these appliances you can skip the
Fuse/L-N Loop Pre-test. To perform the test on
appliances that fail the Fuse/L-N Loop Pre-test do the
following:
release
LOAD/
LEAK
and press it again before the FAIL X
indication is removed from the display, (press as described
at step 2 of the test procedure).