PELV Test
The PELV (Protective Extra Low Voltage) test measures
the voltage on the PROBE PELV input when the idle
screen is being displayed..
To perform the PELV test, do the following:
1
STOP
Revert to the idle screen if it is not already
being displayed.
2
Connect the test probe to the tester PROBE
PELV input and connect the appliance to a
mains supply socket.
3
Apply the test probe to the test point.
The display shows the test result, see the
table below.
4
MEM
Store the test result, if required.
DISPLAY EXPLANATION
> V
AC
The PELV is above the acceptable limit.
V
AC
The PELV is below the acceptable limit,
the display shows the IDLE screen.
Using the Memory
The tester has a non-volatile memory to save a minimum
of 100 test results. The power-on screen shows a
message if the memory is full or nearly full:
memory > : the store is nearly full (>75%)
memory > : the store is completely full
If one of these messages is shown you should print the
stored test results (see page 27), and then clear the store
(see page 27).